日本KOMYOKK乙炔气体检测管101S ,日本KOMYOKK乙炔气体检测管101S ACETYLENE HCCH 50-1000ppm,日本光明理化工业株式会社从1947年开始,在气体分析领域开拓了六十余年 , 形成了驰名于世的 ” 北川 ” 式气体检测管。气体快速分析法 ” 北川 ” 的历史就是日本检测管的历史 。目前是世界上*大的检测管生产厂家,也是日本**家通过美国SEI的认证,欧盟的认证的检测管厂家。公司不仅有ISO9001:2000的认证,还获得了环境管理的ISO14001的认证。北川式手动气体采集器AP-20 备有采气完成指示计 , 可以清楚的判明采气是否完成 . • 有专门截留玻璃的截留槽 , 以防止检测管割尖时玻璃片飞溅 . • 有采气容积为 50 毫升和 100 毫升的两段式制动卡 . • 采气器极为小巧 , 并有便于携带的皮具包 。
Tube No.101S
1. PERFORMANCE
1) Measuring range :50-1000 ppm
Number of pump strokes :1(100mr)
2) Sampling time :3 minutes/1 pump stroke
3) Detectable limit :10 ppm
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :Pale yellow→Brownish blue
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD.-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Molybadate is reduced and molybdeum blue is produced.
HC≡CH+PdSO4+(NH4)2MoO4→Mo3O8
4. CALIBRATION OF THE TUBE
STANDARD GAS CYLINDER METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
100
50
200
300
400
500
600
700
800
900
1000
100ml
ppm
GAS To pump
101S ACETYLENE
Substance Interference ppm Coexistence
Whole layer is discoloured to Whole layer is discoloured to
Carbon Monoxide Green or Blue. 50 Greraedeinng so ra Brel ugeiv eann.d higher
Whole layer is discoloured to
Hydrogen (over 40℃) 〃10% Blue and higher readings are
given.
Unsaturated hydrocarbons Higher readings are given.
such as Ethylene, Propylene, Similar stain is produced. 10
Butylene, etc.
Saturated hydrocarbons such
as Propane, Butane, etc. 〃〃
(more than C3)
Butadiene Original colour is faded to white. 25 Original colour is faded toWhite
and lower readings are given.
Hydrogen sulphide Whole layer is discoloured to Black. 10 Black stain is produced.
Ammonia Original colour is faded to White. Original colour is faded to White
and lower readings are given.
Hydrogen cyanide 〃Blue stain is produced and higher
reading are given.
Chlorine bYreollwown isstha ionr aisn pgreo odru cYeedllo. wish Higher readings are given.
Nitrogen dioxide 〃〃
Carbon disulphide 〃〃
Benzene 〃〃
1. PERFORMANCE
1) Measuring range :50-1000 ppm
Number of pump strokes :1(100mr)
2) Sampling time :3 minutes/1 pump stroke
3) Detectable limit :10 ppm
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :Pale yellow→Brownish blue
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD.-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Molybadate is reduced and molybdeum blue is produced.
HC≡CH+PdSO4+(NH4)2MoO4→Mo3O8
4. CALIBRATION OF THE TUBE
STANDARD GAS CYLINDER METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
100
50
200
300
400
500
600
700
800
900
1000
100ml
ppm
GAS To pump
101S ACETYLENE
Substance Interference ppm Coexistence
Whole layer is discoloured to Whole layer is discoloured to
Carbon Monoxide Green or Blue. 50 Greraedeinng so ra Brel ugeiv eann.d higher
Whole layer is discoloured to
Hydrogen (over 40℃) 〃10% Blue and higher readings are
given.
Unsaturated hydrocarbons Higher readings are given.
such as Ethylene, Propylene, Similar stain is produced. 10
Butylene, etc.
Saturated hydrocarbons such
as Propane, Butane, etc. 〃〃
(more than C3)
Butadiene Original colour is faded to white. 25 Original colour is faded toWhite
and lower readings are given.
Hydrogen sulphide Whole layer is discoloured to Black. 10 Black stain is produced.
Ammonia Original colour is faded to White. Original colour is faded to White
and lower readings are given.
Hydrogen cyanide 〃Blue stain is produced and higher
reading are given.
Chlorine bYreollwown isstha ionr aisn pgreo odru cYeedllo. wish Higher readings are given.
Nitrogen dioxide 〃〃
Carbon disulphide 〃〃
Benzene 〃〃